A label-free immunosensor for sensitive detection of RACK 1 cancer biomarker based on conjugated polymer modified ITO electrode
Özet
A new flexible biosensor based on conjugated polymer functionalized indium tin oxide (ITO) sheet was fabricated for Receptor for Activated C Kinase 1 (RACK 1) determination. Poly(3-thiophene acetic acid) (P(Thi-Ac)) was used as an immobilization matrix for construction of RACK 1 immunosensor. This polymer had a great number of carboxyl groups on its end site and these carboxyl ends provided anchoring points to the anti-RACK 1 antibodies. Anti-RACK 1 antibodies were covalently attached on the ITO electrode and recognized the RACK 1 antigens. Electrochemical characterizations were made by employing electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV) techniques. Additionally, single frequency impedance method (SFI) was applied to follow the specific biointeraction between antibody and antigen. As a result of specific biointeraction, the designed immunosensor exhibited a wide linear detection range between 0.01 pg/mL and 2 pg/mL RACK 1 with a detection limit of 3.1 fg/mL. Scanning electron microscopy and atomic force microscopy analyses were employed for electrode surface morphology investigation. The designed RACK 1 biosensor had good repeatability (5.73 %, RSD), excellent reproducibility (2.5 %, RSD), long storage-stability and reusable property. In addition, the fabricated RACK 1 biosensor was applied to determine RACK 1 concentration in human serums and the recovery was ranging from 98.79%–100.22%. This work illustrated a new tool to construct a sensitive and low-cost disposable biosensor for applications in clinical monitoring. © 2020 Elsevier B.V.
Cilt
190İlgili Öğeler
Başlık, yazar, küratör ve konuya göre gösterilen ilgili öğeler.
-
Fabrication of electrochemical immunosensor based on acid-substituted poly(pyrrole) polymer modified disposable ITO electrode for sensitive detection of CCR4 cancer biomarker in human serum
Aydın, Elif Burcu; Aydın, Muhammet; Sezgintürk, Mustafa Kemal (Elsevier B.V., 2021)This study described the first impedimetric immunosensor reported for the determination of CCR4, a new prostate cancer biomarker. This impedimetric immunosensor was constructed through the modification of disposable indium ... -
A novel electrochemical immunosensor based on acetylene black/epoxy-substituted-polypyrrole polymer composite for the highly sensitive and selective detection of interleukin 6
Aydın, Elif Burcu; Aydın, Muhammet; Sezgintürk, Mustafa Kemal (Elsevier B.V., 2021)An ultrasensitive immunosensor based on acetylene black (AB)/epoxy-substituted-poly(pyrrole) polymer (EpxS-PPyr) composite coated disposable indium tin oxide (ITO) electrode was fabricated for interleukin 6 (IL 6) detection. ... -
A sensitive and selective approach for detection of IL 1? cancer biomarker using disposable ITO electrode modified with epoxy-substituted polythiophene polymer
Aydın, Muhammet (Elsevier Ltd, 2019)A highly sensitive and label-free electrochemical immunosensor for sensitive detection of interleukin 1? cancer biomarker was described by using epoxy-substituted polythiophene polymer modified disposable indium tin oxide ...