Selective and ultrasensitive electrochemical immunosensing of NSE cancer biomarker in human serum using epoxy-substituted poly(pyrrole) polymer modified disposable ITO electrode
Özet
An ultrasensitive, specific and label-free immunosensor was developed to determine the lowest alteration of neuron specific enolase antigen (NSE), the standard biomarker of lung cancer patients. This immunosensor was the first biosensor which was fabricated by utilizing epoxy-substituted-polypyrrole P(Pyr-Epx) polymer as an immobilization platform. The fabrication steps of the immunosensor were followed by using electrochemical studies (electrochemical impedance spectroscopy and cyclic voltammetry) and these techniques were applied to characterize the binding interactions. Scanning electron microscopy and atomic force microscopy analyses were performed to investigate the changes formed on the indium tin oxide electrode surface morphology and electrode surface structure. The optimization of the experimental parameters and the analytical performance of the proposed immunosensor were widely evaluated. Under optimum experimental conditions, the change in impedimetric signal was determined to follow the specific biointeraction between anti-NSE antibody and NSE antigen. The sensing ability of the proposed immunosensor for NSE detection showed a wide linear detection range of 0.02–7.5 pg/mL, with a low detection limit of 6.1 fg/mL. This impedimetric immunosensor also demonstrated highly repeatable and reproducible responses, and reliable results in the analysis of human serum samples with recoveries between 98.29 % and 102.81 %. In addition, the designed immunosensor could be used for detection of lung cancer biomarkers after simple dilution of serum samples with phosphate buffer. © 2019 Elsevier B.V.
Cilt
306İlgili Öğeler
Başlık, yazar, küratör ve konuya göre gösterilen ilgili öğeler.
-
A comparison between LP(GMA) and CLP(GMA) polymer composites as an immobilization matrix for biosensing applications: A model immunosensor for IL 1a
Aydın, Elif Burcu; Sezgintürk, Mustafa Kemal (Elsevier B.V., 2019)A comparative study was made by using two matrices (Linear poly(glycidyl methacrylate) (GMA) polymer (LP(GMA)) and carbon black-polyvinylidene fluoride (PVDF)-LP(GMA) composite CLP(GMA)) as immobilization platforms for ... -
A novel electrochemical immunosensor based on acetylene black/epoxy-substituted-polypyrrole polymer composite for the highly sensitive and selective detection of interleukin 6
Aydın, Elif Burcu; Aydın, Muhammet; Sezgintürk, Mustafa Kemal (Elsevier B.V., 2021)An ultrasensitive immunosensor based on acetylene black (AB)/epoxy-substituted-poly(pyrrole) polymer (EpxS-PPyr) composite coated disposable indium tin oxide (ITO) electrode was fabricated for interleukin 6 (IL 6) detection. ... -
Fabrication of electrochemical immunosensor based on acid-substituted poly(pyrrole) polymer modified disposable ITO electrode for sensitive detection of CCR4 cancer biomarker in human serum
Aydın, Elif Burcu; Aydın, Muhammet; Sezgintürk, Mustafa Kemal (Elsevier B.V., 2021)This study described the first impedimetric immunosensor reported for the determination of CCR4, a new prostate cancer biomarker. This impedimetric immunosensor was constructed through the modification of disposable indium ...