Determination of Optimum Parameters for Cochlear Implants Speech Processors by Using Objective Measures

dc.authorscopusid57203165669
dc.contributor.authorArslan, Özkan
dc.date.accessioned2023-04-20T08:05:54Z
dc.date.available2023-04-20T08:05:54Z
dc.date.issued2022
dc.departmentFakülteler, Çorlu Mühendislik Fakültesi, Elektronik ve Haberleşme Mühendisliği Bölümü
dc.description.abstractIn a cochlear implant (CI) speech processor, several parameters such as channel numbers, bandwidths, rectification type, and cutoff frequency play an important role in acquiring enhanced speech. The effective and general purpose CI approach has been a research topic for a long time. In this study, it is aimed to determine the optimum parameters for CI users by using different channel numbers (4, 8, 12, 16, and 22), rectification types (half and full) and cutoff frequencies (200, 250, 300, 350, and 400 Hz). The CI approaches have been tested on Turkish sentences which are taken from METU database. The optimum CI structure has been tested with objective quality that weighted spectral slope (WSS) and objective intelligibility measures such as short-term objective intelligibility (STOI) and perceptual evaluation of speech quality (PESQ). Experimental results show that 400 Hz cutoff frequency, full wave rectifier, and 16-channels CI approach give better quality and higher intelligibility scores than other CI approaches according to STOI, PESQ and WSS results. The proposed CI approach provides the ability to percept 91% of output vocoded Turkish speech for CI users. © 2022, TUBITAK. All rights reserved.
dc.identifier.doi10.31202/ecjse.1011025
dc.identifier.endpage947
dc.identifier.issn2148-3736
dc.identifier.issue3en_US
dc.identifier.scopus2-s2.0-85139794553
dc.identifier.scopusqualityQ4
dc.identifier.startpage936
dc.identifier.urihttps://doi.org/10.31202/ecjse.1011025
dc.identifier.urihttps://hdl.handle.net/20.500.11776/11087
dc.identifier.volume9
dc.indekslendigikaynakScopus
dc.institutionauthorArslan, Özkan
dc.language.isoen
dc.publisherTUBITAK
dc.relation.ispartofEl-Cezeri Journal of Science and Engineering
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectCochlear implant
dc.subjectfilter bank
dc.subjectobjective intelligibility measures
dc.subjectvocoder
dc.titleDetermination of Optimum Parameters for Cochlear Implants Speech Processors by Using Objective Measures